Remote Control Commands
R&S
®
SMBV100B
843
User Manual 1178.4460.02 ─ 03
Parameters:
<Offset>
float
Range:
depends on lfo voltage
Increment: 0.001
*RST:
0
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce]:LFOutput<ch>:SOURce
<Source>
Determines the LF signal to be synchronized, when monitoring is enabled.
Parameters:
<Source>
LF1 | LF2 | NOISe | AM | FMPM | EXT1
LF1|LF2
Selects an internally generated LF signal.
NOISe
Selects an internally generated noise signal.
EXT1
Selects an externally supplied LF signal
AM
Selects the AM signal.
FMPM
Selects the signal also used by the frequency or phase modula-
tions.
*RST:
LF1
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce]:LFOutput:SIMPedance
<SImpedance>
Selects the impedance of the LF output.
Parameters:
<SImpedance>
G600 | G50
*RST:
G50
Example:
SOURce:LFOutput:SIMPedance G600
Sets the output impedance of the LF generator to 600 Ohm.
Manual operation:
See
[:SOURce]:LFOutput<ch>:SOURce:PATH
<SourPath>
Determines the path of the LF output source.
SOURce Subsystem