Configuring the RF Signal
R&S
®
SMBV100B
389
User Manual 1178.4460.02 ─ 03
6.12.3.3
Fill with Sensor
Access:
1. Select "RF" > "RF Level" > "User Correction".
2. Select "UCOR Data > navigate to the file *.ucor > Select".
3. Select "Edit UCOR Data > Fill With Sensor...".
This dialog contains parameters for filling a table automatically with sensor read-
ings.
To fill the table, select "Execute".
For information on power sensors an how to use them, see
Calibrate the Power Level with an R&S NRP Power Sensor"
Settings
Fill User Correction Data with Sensor
......................................................................... 389
Used SMBV100B Settings For Measurement
.............................................................390
Fill User Correction Data with Sensor
●
"Sensor"
Displays connected sensors for selection.
●
"List To Fill"
Indicates the used list.
●
"Use SParameter"
Indicates whether SParameter correction in the R&S NRP power sensor is used.
●
"Include Zeroing"
Performs a zeroing procedure before acquiring the user correction data to improve
precision.
No signal is applied to the sensor during zeroing. RF output is temporarily switched
off during that time.
When unchecked, the zeroing procedure is skipped. However, the RF signal level
might be blanked shortly. This setting is useful if blanking of RF is undesirable or
the absence of power at the sensor cannot be guaranteed.
●
"Execute"
Improving Level Performance