Remote Control Commands
R&S
®
SMBV100B
804
User Manual 1178.4460.02 ─ 03
Parameters:
<Delay>
float
Range:
0 to 10E-6
Increment: 1E-12
*RST:
0
Example:
SOURce:BB:IMPairment:RF1:DELay 32.0E-9
Manual operation:
See
[:SOURce<hw>]:IQ:IMPairment:IQRatio[:MAGNitude]
<Magnitude>
[:SOURce]:BB:IMPairment:IQOutput<ch>:IQRatio[:MAGNitude]
<IqRatio>
[:SOURce]:BB:IMPairment:RF<ch>:IQRatio[:MAGNitude]
<IqRatio>
Sets the ratio of I modulation to Q modulation (amplification imbalance) of the corre-
sponding digital channel.
Table 12-3: Value range
Impairments
Min [dB]
Max [dB]
Resolution
Digital
-4
4
0.0001
Analog
-1
1
0.0001
Parameters:
<IqRatio>
float
The setting value can be either in dB or %. An input in percent is
rounded to the closest valid value in dB.
Range:
-4 to 4
Increment: 1E-4
*RST:
0
Default unit: dB | PCT (setting command) / dB (result value)
Example:
SOURce:BB:IMPairment:RF1:IQRatio:MAGNitude 10 PCT
SOURce:BB:IMPairment:RF1:IQRatio:MAGNitude?
// 0.848
// the value is returned in dB
Manual operation:
See
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I
<I>
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q
<Q>
[:SOURce]:BB:IMPairment:IQOutput<ch>:LEAKage:I
<I>
[:SOURce]:BB:IMPairment:IQOutput<ch>:LEAKage:Q
<Q>
[:SOURce]:BB:IMPairment:RF<ch>:LEAKage:I
<I>
[:SOURce]:BB:IMPairment:RF<ch>:LEAKage:Q
<Q>
Determines the leakage amplitude of the I or Q signal component of the corresponding
stream
SOURce Subsystem