![Rohde & Schwarz SMBV100B User Manual Download Page 409](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881409.webp)
Monitoring Signal Characteristics
R&S
®
SMBV100B
407
User Manual 1178.4460.02 ─ 03
7
Monitoring Signal Characteristics
The R&S
SMBV100B is a highly configurable signal generator with level settings dis-
tributed over several dialogs. To optimize the signal generation process and to help
monitoring the variation of the signal parameters through the instrument, the
R&S
SMBV100B is equipped with a real time graphical signal display. This section
explains the provided functions for monitoring of the signal characteristics in real time.
Here you can also find out how to visualize the generated signal with measurement
diagrams.
7.1
Displaying Baseband Signal Characteristics in Real-
Time
In test setups, you usually use signal and spectrum analyzers or an oscilloscope to dis-
play and analyze the characteristics of the generated signal. In the R&S
SMBV100B,
you can also use the built-in graphical signal display function and visualize a subset of
signal characteristics.
7.1.1
Required Options
The minimum equipment layout for using this function includes the base unit.
7.1.2
About the Graphical Signal Display
To ease the setting and routing of baseband signals, R&S
SMBV100B is able to cap-
ture these signals and display them graphically in different ways: as I/Q, vector, eye or
constellation diagram, as power spectrum or as statistical evaluations.
In the digital domain, the measurements are based on recorded I/Q samples. The
sources of these samples are selectable points (acquisition points) in the signal flow.
You can configure and display the signal characteristics simultaneously in up to dia-
grams.
7.1.2.1
Available Diagrams
This section focuses on graphical signal displays, their content and application. See
for an overview of the signal characteristics you can monitor.
Table 7-1: Graphics modes overview
Graphics modes
Domain
Measurement is known from
this instrument
Time
Oscilloscope (standard mode)
Time
Oscilloscope (XY mode)
Displaying Baseband Signal Characteristics in Real-Time