Multi-Instrument Setups
R&S
®
SMBV100B
446
User Manual 1178.4460.02 ─ 03
Connect/Disconnect All Remote
Triggers the instrument to:
●
Establish the remote connections to all configured
.
●
To terminate all existing remote connections.
The external instruments are not removed from the list of external instruments.
Only the remote connection status changes.
Remote connection status can be toggled with the parameter
.
Note:
External instruments connected to the R&S
) whenever:
●
The connection to the external instrument is established or terminated ("Connect/
Disconnect All Remote" or "Rem. Conn > On/Off")
●
The R&S
SMBV100B is preset ([Preset]).
To ensure that the remote connections to the external instruments will be automatically
established after instrument's start-up or a reboot, set
Remote command:
:SCONfiguration:EXTernal:REMote:CONNect[:ALL]
:SCONfiguration:EXTernal:REMote:DISConnect[:ALL]
Preset behavior: Keep connections to external instruments
Sets the effect of R&S
SMBV100B preset ([Preset] or
*RST
) on the following:
●
The connection status between the instrument and the connected external instru-
ments.
●
The R&S
SMBV100B itself and the preset values.
●
The connected instruments.
"Preset behavior..."
Off
On
R&S
SMBV100B and preset
values
Standard R&S
SMBV100B preset
.
Standard R&S
SMBV100B preset
Excluded from preset:
●
"I/Q Output Type"
Connection status
Terminated ("Disconnect All Remote")
Unchanged
Connected instruments
External instrument preset (
) External instrument preset
Remote command:
:SCONfiguration:EXTernal:PBEHaviour
Connector Name
Displays the connector name.
Remote command:
via keyword in the SCPI syntax, e.g.
IQOutput
Direction
Displays the connector direction as a color coded field.
System Configuration Settings