Example: Performing a simple power measurement
Prerequisite: The sensor is connected to the instrument and mapped to the first sensor
channel.
:INITiate1:CONTinuous ON
//Switches the continous power measurement on
:READ1?
// Triggers the measurement and displays the results
Example: Performing a power measurement with a fixed filter
Prerequisite: The sensor is connected to the instrument and mapped to the first sensor
channel.
SENSe1:SOURce RF
//Sensor measures the power of the RF signal
SENSe1:FILTer:TYPE NSRatio
//Selects fixed noise filter mode
SENSe1:FILTer:NSRatio 0.02 DB
//Sets the maximum noise component in the result to 0.02 DB
SENSe1:FILTer:NSRatio:MTIMe 10
//Limits the settling time to 10 seconds.
:SENSe1:APERture:DEFault:STATe 0
// Deactivates the default aperture time of the sensor
:SENSe1:APERture:TIMe 10e-6
// Sets the aperture time to 10 us
SENSe1:UNIT DBM
//Selects unit dBm for the measured value
:INITiate:CONTinuous ON
//Switches the continous power measurement on
:READ?
//Triggers the measurement and displays the results
........................................................................................653
:INITiate<hw>[:POWer]:CONTinuous
............................................................................... 653
............................................................................................654
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
.............................................................. 655
:SENSe<ch>[:POWer]:APERture:TIMe
.............................................................................655
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
.......................................................655
SENSe, READ, INITiate and SLISt Subsystems