Configuring the RF Signal
R&S
®
SMBV100B
334
User Manual 1178.4460.02 ─ 03
Settings
.............................................................................................. 334
................................................................................................334
............................................................................................... 334
Start Level / Stop Level
Defines the RF level sweep range by setting the start and end values.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Remote command:
Spacing
Indicates that the level step intervals are linearly (straight proportional) distributed, i.e.
intervals are of equal size.
Set the step size with the parameter "Step".
Remote command:
n.a.
Step
Applies to "RF Level Sweep" mode.
Sets the step width for the RF level sweep steps logarithmically in dB.
Remote command:
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]
Use Digital Attenuation
Activates digital attenuation during RF level sweep.
Setting Characteristics
Selects additional quality characteristics to optimize the behavior of the RF signal level
for the corresponding application.
"Auto"
Sets the RF output level automatically according to the selected
mode.
In this mode, the instrument provides the highest dynamic range and
fastest setting times, as specified in the data sheet.
The RF signal is shortly blanked when the step attenuator is switch-
ing.
Varying the RF Signal in List or Sweep Mode