![Rohde & Schwarz SMBV100B User Manual Download Page 729](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881729.webp)
Remote Control Commands
R&S
®
SMBV100B
727
User Manual 1178.4460.02 ─ 03
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:EXECute
Executes a trigger.
Example:
See
Example "Adjusting clock, marker and trigger settings"
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:DELay:UNIT
<TrigDelUnit>
Determines the units the trigger delay is expressed in.
Parameters:
<TrigDelUnit>
SAMPle | TIME
*RST:
SAMPle
Example:
See
Example "Specifying delay and inhibit values in time units"
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:DELay
<Delay>
Specifies the trigger delay.
Parameters:
<Delay>
float
Range:
0 to 2147483647
Increment: 0.01
*RST:
0
Default unit: symbol
Example:
See
Example "Adjusting clock, marker and trigger settings"
Manual operation:
See
"(Specified) External Delay/(Specified) Trigger Delay"
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:TDELay
<ExtTimeDelay>
Specifies the trigger delay for external triggering. The value affects all external trigger
signals.
SOURce Subsystem