![Rohde & Schwarz SMBV100B User Manual Download Page 41](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881041.webp)
Getting Started
R&S
®
SMBV100B
39
User Manual 1178.4460.02 ─ 03
●
<Serial Number>
is the individual serial number of the instrument.
You can find the serial number at the rear panel of instrument. It is the third part of
the device ID printed on the barcode sticker .
Example:
The default hostname of an R&S
SMBV100B with a serial number 102030 is
SMBV100B-102030
.
To query and change a computer name
1. Press the [Setup] key.
2. Select "Remote Access > Network".
The computer name is displayed under "Hostname".
3. Press the [Setup] key.
4. Select "Security > Protection".
5. Enable the "Protection Level 1".
The default password is
123456
.
The parameter "Hostname" in the "Network" tab is now enabled for configuration.
6. Change the "Hostname".
2.1.4.3
Assigning the IP Address
Depending on the network capacities, the TCP/IP address information for the instru-
ment can be obtained in different ways.
●
If the network supports DHCP (dynamic host configuration protocol), the address
information is assigned automatically.
●
If the network does not support DHCP, the instrument tries to obtain the IP address
via Zeroconf (APIPA) protocol. If this attempt does not succeed or if the instrument
is set to use alternate TCP/IP configuration, the addresses must be set manually.
Since the dynamic TCP/IP configuration assigns the address information automatically,
it is safe to establish a physical connection to the LAN without any previous instrument
configuration.
Preparing for Use