Remote Control Commands
R&S
®
SMBV100B
630
User Manual 1178.4460.02 ─ 03
Example:
CAL:ALL:MEAS?
0
Executes the adjustments of all instrument functions. When
completed, it indicates that the adjustment has been performed
successfully.
Usage:
Query only
Manual operation:
See
:CALibration<hw>:ALL:DATE?
Queries the date of the most recently executed full adjustment.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Date>
string
Example:
CAL:ALL:DATE?
// "2016-01-01"
Usage:
Query only
Manual operation:
See
:CALibration<hw>:ALL:INFormation?
Queries the current state of the internal adjustment.
Return values:
<CalInfoText>
string
Example:
CAL:ALL:INF?
"Instrument is calibrated, no adjustment required."
"UNCAL, instrument is warming up."
"UNCAL, Please perform full adjustment after warming up."
"UNCAL, Please perform full adjustment."
Usage:
Query only
Manual operation:
See
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Temperature>
string
CALibration Subsystem