Configuring the RF Signal
R&S
®
SMBV100B
330
User Manual 1178.4460.02 ─ 03
"Triangle"
The sweep runs from start to stop value and back, i.e. the shape of
the sweep resembles a triangle. Each subsequent sweep starts at the
start frequency.
Figure 6-17: Sweep signal triangle shape
1 = Start value
2 = Stop value
3 = Step width
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe
[:SOURce<hw>]:SWEep:POWer:SHAPe
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SHAPe
Spacing
In "RF and LF Frequency Sweep" modes, selects the mode for the calculation of the
frequency intervals, with which the current frequency at each step is increased or
decreased.
Enter the step size with the parameter
"Linear"
Takes the frequency value entered as absolute value in Hz.
"Logarithmic"
Takes the value entered as a logarithmic value, i.e. as a constant
fraction of the current frequency in %.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing
Dwell Time
Defines the duration of the individual sweep steps.
See also
Chapter 6.10.4, "Significant Parameters and Functions"
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
[:SOURce<hw>]:SWEep:POWer:DWELl
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl
Varying the RF Signal in List or Sweep Mode