![Rohde & Schwarz SMBV100B User Manual Download Page 824](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881824.webp)
Remote Control Commands
R&S
®
SMBV100B
822
User Manual 1178.4460.02 ─ 03
// ******************************************************************
// If the attenuation between the interconnection LO OUT and LO IN is
// greater than 1 dB at the oscillator frequency,
// we recommend that you perform the LO level adjustment
// ******************************************************************
:CALibration:LOSCillator:Coupling:LOCAL?
// ******************************************************************
// Address and reset the third instrument
// ******************************************************************
*RST; *CLS
// ******************************************************************
// Configure the oscillator signal of the third next instrument
// ******************************************************************
SOURce:FREQuency 5 GHz
SOURce:FREQuency:LOSCillator:MODe EXTernal
SOURce:FREQuency:LOSCillator:OUTPut:STATe OFF
See also:
●
"To configure LO coupling with cascaded instruments to achieve phase coherence"
[:SOURce<hw>]:FREQuency:LOSCillator:INPut:FREQuency?
........................................... 822
[:SOURce<hw>]:FREQuency:LOSCillator:MODE
.............................................................. 823
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:FREQuency?
........................................823
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe
.................................................. 823
................................................................................824
[:SOURce<hw>]:FREQuency[:CW|FIXed]
.........................................................................824
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL
................................................................. 825
[:SOURce<hw>]:FREQuency:MANual
..............................................................................825
[:SOURce<hw>]:FREQuency:MULTiplier
.......................................................................... 826
[:SOURce<hw>]:FREQuency:OFFSet
.............................................................................. 826
[:SOURce<hw>]:FREQuency:CENTer
.............................................................................. 827
.................................................................................827
................................................................................ 827
.................................................................................828
[:SOURce<hw>]:FREQuency:STEP:MODE
...................................................................... 828
[:SOURce<hw>]:FREQuency:STEP[:INCRement]
............................................................. 828
[:SOURce<hw>]:FREQuency:LOSCillator:INPut:FREQuency?
Queries the required external reference frequency.
Return values:
<Frequency>
float
Range:
100E3 to 20E9
Increment: 0.01
*RST:
0
Example:
See
Example "Set up of LO coupling and LO level adjustment
SOURce Subsystem