![Rohde & Schwarz SMBV100B User Manual Download Page 887](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881887.webp)
Remote Control Commands
R&S
®
SMBV100B
885
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the level sweep spacing. The sweep spacing for level sweeps is always linear.
Return values:
<Mode>
LINear
*RST:
LINear
Example:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result:
"LIN"
linear spacing
Usage:
Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]
<Logarithmic>
Sets a logarithmically determined step size for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Logarithmic>
float
The unit
dB
is mandatory.
Range:
0.01 to 139 dB
Increment: 0.01
*RST:
1
Default unit: dB
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
<Dwell>
Sets the dwell time for a frequency sweep step.
Parameters:
<Dwell>
float
Range:
0.001 to 100
Increment: 100E-6
*RST:
0.01
Manual operation:
See
SOURce Subsystem