Remote Control Commands
R&S
®
SMBV100B
878
User Manual 1178.4460.02 ─ 03
[:SOURce]:ROSCillator:PRESet
Resets the reference oscillator settings.
Example:
See
Example "Configuring the reference oscillator"
Usage:
Event
Manual operation:
See
[:SOURce]:ROSCillator:SOURce
<Source>
Selects between internal or external reference frequency.
Parameters:
<Source>
INTernal | EXTernal
*RST:
n.a. (factory preset: INTernal)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
<State>
Determines that the RF output is turned off when the external reference signal is
selected, but missing.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
"Deactivate RF Output (if external reference is missing)"
[:SOURce]:ROSCillator:EXTernal:FREQuency
<Frequency>
Sets the frequency of the external reference.
Parameters:
<Frequency>
100MHZ | 1GHZ | VARiable | 10MHZ
*RST:
n.a. (factory preset: 10MHZ)
Example:
See
Example "Configuring the reference oscillator"
Options:
VARiable requires R&S
SMBVB-K704
1GHZ requires R&S
SMBVB-B3
Manual operation:
See
"External Reference Frequency"
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable
<Frequency>
Specifies the user-defined external reference frequency.
SOURce Subsystem