![Rohde & Schwarz SMBV100B User Manual Download Page 810](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881810.webp)
Remote Control Commands
R&S
®
SMBV100B
808
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:GRAPhics:SRATe:USER
<User>
(Enabled for
BB:GRAPH:SRAT:MODE USER
)
Selects the signal bandwidth for the diagram. The setting range moves between the
minimum and maximum bandwidth which is possible for the selected graphical signal
display. The selection is made graphically by moving the pointer.
Parameters:
<User>
float
Range:
0.01 to 100
Increment: 0.01
*RST:
10
Default unit: PCT
Example:
BB:GRAP:SRAT:USER 20
sets the sample rate factor.
Example:
BB:GRAP:SRAT:MODE USER
sets the sample rate mode.
Manual operation:
See
[:SOURce]:BB:GRAPhics:CLOSe
Closes all graphical signal displays.
Usage:
Event
Manual operation:
See
[:SOURce]:BB:GRAPhics:ADD
<Size>
Adds a graphical signal display (according to the current
MODE
,
SOURce
,
SRATe:*
and
TRIGger:*
settings).
Setting parameters:
<Size>
MAXimized | MINimized
Usage:
Setting only
Manual operation:
See
[:SOURce]:BB:GRAPhics:SOURce
<Source>
Defines the signal acquisition point, that is the location in the signal flow where the dis-
played signal is tapped from.
See
Chapter 7.1.2.2, "Signal Acquisition Points"
Parameters:
<Source>
STRA | BBA | RFA | BBIA | IQO1
STRA
Stream A; input stream of the "IQ Stream Mapper"
SOURce Subsystem