![Rohde & Schwarz SMBV100B User Manual Download Page 842](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881842.webp)
Remote Control Commands
R&S
®
SMBV100B
840
User Manual 1178.4460.02 ─ 03
[:SOURce]:LFOutput<ch>:BANDwidth?
Queries the bandwidth of the external LF signal.
Return values:
<Bandwidth>
BW0M2 | BW10m
*RST:
BW10m
Example:
LFO:BAND?
// BW10m
// the bandwidth of the externally supplied LF signal is 10 MHz
Usage:
Query only
Manual operation:
See
[:SOURce]:LFOutput<ch>:FREQuency
<Frequency>
Sets the frequency of the LF signal in
[:SOURce<hw>]:LFOutput:FREQuency:
CW|FIXed
mode.
Note
●
If signal source "Internal" is set, the instrument performs the analog modulations
(AM/FM/PhiM/PM) with this frequency.
●
In sweep mode (
[:SOURce<hw>]:LFOutput:FREQuency:MODE
SWE
), the fre-
quency is coupled with the sweep frequency.
Parameters:
<Frequency>
float
Range:
depends on the installed options
*RST:
1000
Default unit: Hz
Example:
SOURce1:LFOutput1:FREQuency 5 kHz
// sets the LF frequency
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:PERiod?
Queries the repetition frequency of the sine signal.
Return values:
<LfSinePeriod>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
0.001
Default unit: s
Example:
See
Example "Configuring the LF generator"
Usage:
Query only
SOURce Subsystem