![Rohde & Schwarz SMBV100B User Manual Download Page 281](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881281.webp)
Applying I/Q Vector Modulation
R&S
®
SMBV100B
279
User Manual 1178.4460.02 ─ 03
System error correction of the I/Q modulator permits precise and repeatable measure-
ments.
If a temperature fluctuation of several degrees occurs, trigger the correction routine
using the following:
●
Select "I/Q Mod > I/Q Modulator > Adjust I/Q Modulator Current Frequency"
Settings:
Adjust I/Q Modulator Current Frequency
.................................................................... 279
State
Activates/deactivates I/Q modulation.
Note:
Interdependency. If "Source > Internal Baseband", the I/Q modulator is always
activated automatically when signal generation is activated ("Baseband > State > On").
I/Q modulation can, however, be deactivated later, e.g. if the relevant signal is the
baseband signal and it is directly routed to the I/Q outputs.
Remote command:
Source
Selects the input signal for the I/Q modulator. Refer to
possible input signals and the connectors they are supplied at.
"Internal Baseband"
Selects the internal baseband signal.
"Analog Wideband I/Q Input"
Selects an external analog signal.
Remote command:
Adjust I/Q Modulator Current Frequency
Starts the adjustment for the I/Q modulator for the currently set frequency and base-
band gain. The I/Q modulator is adjusted regarding carrier leakage, I/Q imbalance and
quadrature.
See also
"To trigger the internal correction routine of the I/Q modulator "
Remote command:
:CALibration<hw>:IQModulator:LOCal?
I/Q Modulator Settings