Remote Control Commands
R&S
®
SMBV100B
770
User Manual 1178.4460.02 ─ 03
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt
<Start>
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP
<Stop>
Selects the last carrier in the carrier range to which the settings shall apply.
Parameters:
<Stop>
integer
Range:
0 to 511
*RST:
0
Example:
BB:ARB:MCAR:EDIT:CARR:STOP 4
The carrier range stops at carrier 4.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
<Step>
Sets the step width by which the starting power of the carriers in the defined carrier
range is incremented.
Parameters:
<Step>
float
Range:
-80 to 80
Increment: 0.01
*RST:
0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -80dB
Sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
The power is incremented by 1dB for each carrier.
That is, the first carrier has -80 dB, the second -79 dB, and so
on.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt]
<Start>
Sets the power for the individual carriers in the defined carrier range.
Parameters:
<Start>
float
Range:
-80 to 0
Increment: 0.01
*RST:
0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -50 dB
sets the power of the carriers in the carrier range to -50 dB.