Getting Started
R&S
®
SMBV100B
50
User Manual 1178.4460.02 ─ 03
I/Q
I and Q inputs for external analog modulation signal directly fed into the I/Q modulator.
For more information, see
Chapter 8.4.1.4, "Overview of the Input and Output Signals
Ref In/Ref Out
Input/output for external reference signal.
Incl. dedicated connectors for the 1GHz reference signal.
See
Chapter 6.8, "Reference Oscillator"
LF
Output for internal LF generator signal.
See also data sheet and user manual, section "Analog Modulation".
Mod Ext
Input for external analog modulation signals.
See
Chapter 6.11, "Analog Modulations"
Pulse Video
Output of the internal pulse generator signal or the looped through pulse signal from
Pulse Ext connector (video signal).
See
Chapter 6.11.4.1, "Pulse Modulation Settings"
Pulse Ext
Input for an external pulse modulation signal or an external trigger/gate signal for the
pulse generator.
Signal Valid
Output signal that marks the valid signal times (valid level and frequency) for all analog
modulations.
Inst Trig
Input for external trigger of sweeps and list mode.
For detailed information on the sweep modes and the triggering, see
"Signal Generation and Triggering in the Sweep and List Modes"
IEC 625/IEEE 488
GPIB-bus interface for remote control of the instrument.
The IEC 625 interface is in compliance with IEEE488 and SCPI. A computer for remote
control can be connected via this interface. To set up the connection, we recommend
that you use a shielded cable.
Note:
To avoid electromagnetic interference (EMI) caused by open lines, always termi-
nate any connected IEC-bus cable with an instrument or a controller.
See also
Chapter B.1, "GPIB-Bus Interface"
Instrument Tour