Configuring the RF Signal
R&S
®
SMBV100B
333
User Manual 1178.4460.02 ─ 03
"Logarithmic"
Takes the value entered as a logarithmic value, i.e. as a constant
fraction of the current frequency in %.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing
Step Linear/Step Logarithmic
In "RF/LF Frequency Sweep" mode, sets the step width for the individual frequency
sweep steps. The value is added at each sweep step to the current frequency.
Depending on the current
, you can enter either an absolute or logarithmic step
width.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
"Step Linear"
The step width is a constant value in Hz.
"Step Logarithmic"
The step width is determined logarithmically in %, i.e. as a constant
fraction of the current frequency.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
6.10.5.3
Level Range Settings
Access:
1. Select "RF" > "Sweep/List" > "RF Level Sweep"
2. Select "Level Range".
Varying the RF Signal in List or Sweep Mode