Remote Control Commands
R&S
®
SMBV100B
888
User Manual 1178.4460.02 ─ 03
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:SWEep:POWer:RUNNing?
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?
Queries the current sweep state.
Return values:
<State>
0 | 1 | OFF | ON
Example:
See
Example "Setup an RF frequency or power sweep"
Usage:
Query only
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
<Logarithmic>
Sets a logarithmically determined step width for the RF frequency sweep. The value is
added at each sweep step to the current frequency.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Logarithmic>
float
The unit is mandatory.
Range:
0.01 to 100
Increment: 1E-3
*RST:
1
Default unit: PCT
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
" Step Linear/Step Logarithmic "
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
<Linear>
Sets the step width for linear sweeps.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Omit the optional keywords so that the command is SCPI-compliant.
SOURce Subsystem