Configuring the RF Signal
R&S
®
SMBV100B
331
User Manual 1178.4460.02 ─ 03
Trigger Slope
For "Mode = Extern Step/Single", selects the polarity of the active slope of an applied
instrument trigger.
Trigger signal is expected at the Inst Trig connector.
"Positive"
The rising edge of the trigger signal triggers the instrument.
"Negative"
The falling edge of the trigger signal triggers the instrument.
Remote command:
Execute Single Sweep
In "Mode = Single", starts a sweep manually.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
[:SOURce<hw>]:SWEep:POWer:EXECute
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute
:TRIGger<hw>[:SWEep][:IMMediate]
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Reset Sweep
Resets a sweep.
With the next trigger event, the sweep starts at the initial value.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL]
6.10.5.2
Frequency Range Settings
Access:
1. Select for example "RF" > "Sweep/List" > "RF Frequency Sweep"
2. Select "Frequency Range".
Varying the RF Signal in List or Sweep Mode