The Spectrum Application
FPC1000
55
User Manual 1178.4130.02 ─ 02
Frequency domain measurements
In the frequency domain, the R&S
FPC1000 analyzes the input signal characteristics
. You can use it, for example, to obtain basic measure-
ment results like peak levels and the shape of the spectrum.
The horizontal axis shows the frequency, the vertical axis shows the signal levels.
The measurement in the frequency domain is the default measurement mode.
Time domain measurements
In the time domain, the R&S
FPC1000 analyzes the characteristics of a signal at a par-
ticular frequency over time. The span during time domain measurements is 0 (
). You can use time domain measurements, for example to monitor the
characteristics of a signal and its behavior over time.
Figure 9-3: Display of a signal in the time domain
The horizontal axis shows the time, the vertical axis shows the signal levels.
You have to select the zero span mode deliberately to measure signal characteristics
in the time domain.
Advanced measurement functions
In addition to these basic measurement functions, you can equip the R&S
FPC1000
with optional advanced measurements.
●
Measuring the Channel Power of Continuously Modulated Signals
Measuring the Occupied Bandwidth
....................................................................... 59
Working with the Spectrogram Result Display
........................................................63
9.2.1 Measuring the Channel Power of Continuously Modulated Signals
Access
(measurement menu): "Meas" > "Measurement Mode" > "Channel Power"
Measurements and Result Displays