Remote Control
FPC1000
190
User Manual 1178.4130.02 ─ 02
[SENSe:]FREQuency:OFFSet
<FrequencyOffset>
This command defines a frequency offset.
Parameters:
<FrequencyOffset>
<numeric value>
Range:
depends on the start frequency to 100 GHz
*RST:
0 Hz
Default unit: Hz
Example:
//Define a frequency offset
FREQ:OFFS 100MHZ
[SENSe:]FREQuency:SPAN
<Span>
This command defines the frequency span.
If you define a span of 0 Hz, the R&S
FPC1000 starts a measurement in the time
domain.
Parameters:
<Span>
<numeric value>
Refer to the datasheet for details on the frequency range of your
R&S
FPC1000.
*RST:
fmax with fmax = maximum frequency
Default unit: Hz
Example:
//Define frequency span
FREQ:SPAN 10MHZ
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Refer to the datasheet for details on the frequency range of your R&S
FPC1000.
Example:
//Restore full span
FREQ:SPAN:FULL
Usage:
Event
[SENSe:]FREQuency:STARt
<Frequency>
This command defines the start frequency for measurements in the frequency domain
(span > 0).
Parameters:
<Frequency>
<numeric value>
Refer to the datasheet for details on the frequency range of your
R&S
FPC1000.
*RST:
0 Hz
Default unit: Hz
Remote Commands of the Spectrum Application