The Spectrum Application
FPC1000
105
User Manual 1178.4130.02 ─ 02
If you enter a positive offset, the two temporary markers are placed below the active
reference point. If you enter a negative value (for example for notch filter measure-
ments), the two temporary markers are placed above the active reference point.
Measuring the signal bandwidth
1. Press the "Mkr" key to open the marker menu.
2. Select the "Marker Mode" menu item to open the marker function menu.
3. Select the "N dB Down" menu item.
The R&S
FPC1000 shows two vertical lines in the diagram that represent the tem-
porary markers and measures the distance (bandwidth) between those two lines.
4. Select the "N dB <x>" menu item to select the distance between the temporary
markers.
Note:
The distance to the reference marker is by default 3
dB below the reference
marker.
9.10.4 Marker Demodulation
Access
: "Mkr" > "Marker Demodulation"
The R&S
FPC1000 features an AM and FM demodulator to demodulate and monitor
audio signals. It demodulates the signal at the marker frequency.
You can listen to the demodulated signal with the internal speaker or headphones that
you can connect to the 3.5
mm headphone jack on the front of the R&S
FPC1000.
When demodulating an AM modulated signal, the R&S
FPC1000 turns the video volt-
age into an audible sound. You should therefore set the reference level to about the
level of the signal that you are demodulating.
If you perform measurements in the time domain, the R&S
FPC1000 demodulates con-
tinuously. In the frequency domain, you can define a time period the R&S
FPC1000
demodulates the signal at the marker frequency. The frequency sweep then stops at
the marker frequency for that time before finishing the sweep.
Remote commands required to configure marker properties:
●
Selecting the demodulation type
1. Press the "Mkr" key to open the marker menu.
2. Select the "Marker Demodulation" menu item to open the marker demodulation
menu.
3. Select the "AM" or the "FM" menu item
The R&S
FPC1000 demodulates the signal according to the demodulation scheme
you have selected.