![Rohde & Schwarz FPC1000 User Manual Download Page 145](http://html.mh-extra.com/html/rohde-and-schwarz/fpc1000/fpc1000_user-manual_1477825145.webp)
Analog Demodulation
FPC1000
130
User Manual 1178.4130.02 ─ 02
Demodulation bandwidth
Modulating frequency [Hz]
Minimum
Maximum
500000
45
7576
300000
30
4545
200000
20
3030
100000
10
1515
50000
5
758
30000
3
455
20000
3
303
10000
3
152
Table 11-6: Minimum and maximum modulating frequencies for audio lowpass: DBW/100
Demodulation bandwidth
Modulating frequency [Hz]
Minimum
Maximum
2000000
60
10000
989000
30
5000
650000
25
3250
500000
15
2500
300000
10
1500
200000
6
1000
100000
3
500
50000
3
250
30000
3
150
20000
3
100
10000
3
50
11.5 Sweep Configuration
Access
: "Sweep"
Remote commands to control the measurement:
●
Chapter 13.12.5, "Sweep Control"
The sweep menu provides settings to control the measurement and the way data is
captured.
Selecting the measurement mode
The measurement mode defines how often the data is demodulated.
Sweep Configuration
北京海洋兴业科技股份有限公司(证券代码:839145)
www.hyxyyq.com