Remote Control
FPC1000
166
User Manual 1178.4130.02 ─ 02
The number of values depends on the length of the queue.
If the error queue is empty, the command returns 0, 'No error'.
Return values:
<ErrorNumber>
Number that contains information about the error. Negative num-
ber: error as defined in the SCPI standard. Positive number:
error that is specific to the R&S
FPC1000.
<ErrorDescription>
String containing a short error description.
Example:
//Query complete error queue
SYST:ERR:ALL?
Usage:
Query only
SYSTem:ERRor:CODE:ALL?
This command queries the complete error queue.
The number of values depends on the length of the queue.
If the error queue is empty, the command returns 0, 'No error'.
Return values:
<ErrorNumber>
Number that contains information about the error. Negative num-
ber: error as defined in the SCPI standard. Positive number:
error that is specific to the R&S
FPC1000.
Example:
//Query complete error queue
SYST:ERR:CODE:ALL?
Usage:
Query only
SYSTem:ERRor:CODE[:NEXT]?
This command queries the code of the next error in the error queue.
If the error queue is empty, the command returns 0.
Return values:
<ErrorNumber>
Number that contains information about the error. Negative num-
ber: error as defined in the SCPI standard. Positive number:
error that is specific to the R&S
FPC1000.
Example:
//Query the code of the next error
STAT:ERR:CODE?
Usage:
Query only
SYSTem:ERRor:COUNt?
This command queries the number of errors currently in the error queue.
Return values:
<NumErrors>
<numeric value>
System Configuration