Remote Control
FPC1000
208
User Manual 1178.4130.02 ─ 02
Return values:
<Data>
<numeric value>
The R&S
FPC1000 returns 1183 values. Each value corre-
sponds to one pixel of a trace.
The unit depends on the measurement and the unit you have
set with
Note:
If you use the auto peak detector, the command reads out
positive peak values only.
Example:
//Read out the data for trace 1.
TRAC:DATA? TRACE1
Usage:
Query only
13.10.8 Marker
CALCulate<n>:DELTamarker<m>:AOFF
..........................................................................208
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT
.......................................................... 209
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK]
........................................................ 209
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK]
......................................................... 209
CALCulate<n>:DELTamarker<m>[:STATe]
.......................................................................209
................................................................................ 210
CALCulate<n>:DELTamarker<m>:X:RELative
.................................................................. 210
CALCulate<n>:DELTamarker<m>:Y?
...............................................................................211
..................................................................................212
CALCulate<n>:MARKer<m>:FREQuency:MODE
.............................................................. 212
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
................................................................212
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
...........................................................212
CALCulate<n>:MARKer<m>:MAXimum:NEXT
..................................................................213
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
................................................................ 213
CALCulate<n>:MARKer<m>:MINimum[:PEAK]
................................................................. 213
CALCulate<n>:MARKer<m>[:STATe]
...............................................................................214
........................................................................................ 214
CALCulate<n>:MARKer<m>:X:SLIMits<slim>:LEFT
.......................................................... 214
CALCulate<n>:MARKer<m>:X:SLIMits<slim>:RIGHt
......................................................... 215
CALCulate<n>:MARKer<m>:X:SLIMits<slim>[:STATe]
...................................................... 216
.......................................................................................216
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off all active delta markers.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Example:
//Turn off all delta markers
CALC:DELT:AOFF
Usage:
Event
Remote Commands of the Spectrum Application
北京海洋兴业科技股份有限公司(证券代码:839145)