Preparing for Use
FPC1000
10
User Manual 1178.4130.02 ─ 02
Risk of instrument damage during operation
An unsuitable operating site or test setup can damage the instrument and connected
devices. Ensure the following operating conditions before you switch on the instrument:
●
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are correctly connected and are not overloaded.
EMI impact on measurement results
Electromagnetic interference (EMI) may affect the measurement results.
To suppress generated electromagnetic interference (EMI):
●
Use suitable shielded cables of high quality. For example, use double-shielded RF
and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
2.1 Unpacking and Checking the Instrument
Unpack the R&S
FPC1000 carefully and check the contents of the package.
●
Check if all items listed on the delivery note, including the getting started manual,
are included in the delivery.
●
Check the R&S
FPC1000 for any damage.
If the contents are damaged, immediately contact the carrier who delivered the
package.
●
Keep the box and packing material.
Unpacking and Checking the Instrument