Remote Control
FPC1000
215
User Manual 1178.4130.02 ─ 02
Prerequisites for this command
●
Turn on marker search limits (
If you perform a measurement in zero span, this command limits the evaluation range
on the trace.
Suffix:
<n>, <m>, <slim>
.
irrelevant
Parameters:
<SearchLimit>
<numeric value>
The unit depends on the measurement: Hz for frequency
domain, seconds for time domain.
Range:
The value range corresponds to the maximum
span.
*RST:
- (is set to the left diagram border when switching
on search limits)
Example:
See
CALCulate<n>:MARKer<m>:X:SLIMits<slim>:RIGHt
CALCulate<n>:MARKer<m>:X:SLIMits<slim>:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range.
Prerequisites for this command
●
Turn on marker search limits (
If you perform a measurement in zero span, this command limits the evaluation range
on the trace.
Suffix:
<n>, <m>, <slim>
.
irrelevant
Parameters:
<SearchLimit>
<numeric value>
The unit depends on the measurement: Hz for frequency
domain, seconds for time domain.
Range:
The value range corresponds to the maximum
span.
*RST:
- (is set to the left diagram border when switching
on search limits)
Example:
//Turn search limits on and define a search range from 10 MHz
to 100 MHz.
CALC:MARK:X:SLIM ON
CALC:MARK:X:SLIM:LEFT 10MHz
CALC:MARK:X:SLIM:RIGH 100MHz