Measurement Wizard
FPC1000
49
User Manual 1178.4130.02 ─ 02
6. Start the measurement sequence with the "Start Measurement" menu item.
The R&S
FPC1000 starts to go through the measurements that are part of the
measurement set. The sequence of the measurement is as defined with the
R&S
InstrumentView software.
Note that it is not possible to change any measurement parameters while using the
measurement wizard. Marker functionality and scaling parameters are, however,
available.
Before each measurement, the wizard shows a message box that contains the
for that measurement.
Figure 8-5: Prepare the measurement as described when this message shows up
7. Prepare the measurements as instructed. When you are done, you can start the
measurement.
8. You have several options before you start a measurement.
● "Continue"
Finishes the current measurement and begins with the next measurement by
showing the necessary preparations.
● "Repeat Measurement"
Repeats the current measurement. Repeating a measurement can be useful,
for example, if the results do not match your expectations and you want to vali-
date the results.
● "Interrupt"
Interrupts the sequence of measurements. Interrupting a measurement can
become necessary if the measurement does not yield the expected results (for
example violated limit lines).
In that case, you can interrupt the measurement sequence and try to find the
origin of the problem by using different settings or measurements than those
defined in the wizard.
When you interrupt the wizard sequence, the complete functionality is available
as if you would not use the wizard. Results of finished measurements remain in
the memory.
When you are finished reconfiguring the measurement, enter the measurement
wizard and resume the measurement sequence with the "Resume Sequence"
menu item.
● "Skip"
Skips a single measurement and initiates the subsequent measurement.
Skipping individual measurements is possible when you turn on "Allow to skip
measurements and finish wizard sequence" in the "Wizard Set" editor of the
R&S
InstrumentView software.
● "Finish Wizard"
Ends the measurement sequence. The results of finished measurements
remain in the memory.