Remote Control
FPC1000
203
User Manual 1178.4130.02 ─ 02
Parameters:
<TriggerSource>
IMMediate
No trigger is applied.
EXTernal
Select an external trigger.
VIDeo
Select the video trigger.
The video trigger is only available if span = 0 (
*RST:
IMMediate
Example:
//Select external trigger
TRIG:SOUR EXT
13.10.7 Traces
CALCulate<n>:MATH<t>:COPY:MEMory
......................................................................... 203
CALCulate<n>:MATH<t>[:EXPRession][:DEFine]
..............................................................204
..................................................................................... 204
DISPlay<n>[:WINDow]:TRACe<t>:MEMory[:STATe]
..........................................................204
DISPlay<n>[:WINDow]:TRACe<t>:MODE
.........................................................................205
DISPlay<n>[:WINDow]:TRACe<t>[:STATe]
.......................................................................205
[SENSe:]DETector<t>[:FUNCtion]
....................................................................................206
[SENSe:]DETector<t>[:FUNCtion]:AUTO
..........................................................................207
.................................................................................................207
CALCulate<n>:MATH<t>:COPY:MEMory
[<MemorySlot>]
This command copies a trace into the memory of the R&S
FPC1000.
Prerequisites for this command
●
Turn on a trace with a trace mode other than "Blank" (
Suffix:
<n>
.
irrelevant
<t>
Setting parameters:
[<MemorySlot>]
not supported
Example:
//Copy the trace into the memory
DISP:TRAC2 ON
DISP:TRAC2:MODE AVER
CALC:MATH2:COPY:MEM
Usage:
Setting only
Remote Commands of the Spectrum Application