Remote Control
FPC1000
232
User Manual 1178.4130.02 ─ 02
DISPlay<n>[:WINDow]:TRACe<t>:Y:SPACing
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:ADJust
●
Reference position:
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RPOSition
●
Reference offset:
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
●
Unit:
●
Preamplifier:
●
Attenuation:
INPut:ATTenuation
INPut:ATTenuation:AUTO
INPut:ATTenuation:MODE
●
Impedance:
INPut:IMPedance
INPut:IMPedance:PAD
●
Transducer:
13.11.4 Bandwidth
Commands to configure the bandwidth described elsewhere.
●
Resolution bandwidth (3 dB):
[SENSe:]BANDwidth[:RESolution]
[SENSe:]BANDwidth[:RESolution]:CISPr
..........................................................................232
[SENSe:]BANDwidth[:RESolution]:CISPr:AUTO
................................................................ 233
[SENSe:]BANDwidth[:RESolution]:CISPr
<Bandwidth>
This command selects the CISPR (6 dB) bandwidth for CISPR conform measure-
ments.
Parameters:
<Bandwidth>
200HZ | 9KHZ | 120KHZ | 1MHZ
Default unit: Hz
Example:
//Select CISPT bandwidth
BAND:CISP 9KHZ