Remote Control
FPC1000
199
User Manual 1178.4130.02 ─ 02
[SENSe:]BWIDth:VIDeo:AUTO
<State>
[SENSe:]BANDwidth:VIDeo:AUTO
<State>
This command turns automatic selection of the resolution bandwidth on and off.
Effects of this command
●
Turn on automatic VBW selection to couple the VBW to the RBW.
When you turn off automatic selection of the video bandwidth, you can select one man-
ually with
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON | 1
Example:
//Select VBW manually
BAND:VID:AUTO OFF
BAND:VID 10HZ
13.10.5 Sweep Control
......................................................................................... 200
ABORt
This command aborts a measurement and resets the trigger system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the
*OPC?
or
*WAI
command after
ABORt
and before the next command.
Note on blocked remote control programs
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S
FPC1000 is blocked for further commands. In this case, you must
interrupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S
FPC1000 on a parallel channel to clear all currently active remote channels.
Depending on the used interface and protocol, send the following commands
●
Visa
:
viClear()
Now you can send the
ABORt
command on the remote channel performing the mea-
surement.
Remote Commands of the Spectrum Application