Remote Control
FPC1000
231
User Manual 1178.4130.02 ─ 02
Parameters:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Define scan range with measurement every 10 MHz
FREQ:MODE SCAN
SCAN:STAR 30MHZ
SCAN:STOP 100MHZ
SCAN:STEP 10MHZ
[SENSe:]SCAN:STEP
<StepSize>
This command defines the frequency step size for a frequency scan.
Prerequisites for this command
●
).
Parameters:
<StepSize>
<numeric value>
Default unit: Hz
Example:
//Define scan range with measurement every 10 MHz
FREQ:MODE SCAN
SCAN:STAR 30MHZ
SCAN:STOP 100MHZ
SCAN:STEP 10MHZ
[SENSe:]SCAN:STOP
<Frequency>
This command defines the start frequency for a frequency scan.
Prerequisites for this command
●
).
Parameters:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Define scan range with measurement every 10 MHz
FREQ:MODE SCAN
SCAN:STAR 30MHZ
SCAN:STOP 100MHZ
SCAN:STEP 10MHZ
13.11.3 Amplitude
Commands to configure amplitude described elsewhere.
●
Reference level:
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel
●
Level range:
Remote Commands of the Receiver Application