Remote Control
FPC1000
226
User Manual 1178.4130.02 ─ 02
Return values:
<Status>
•
0 for PASS
•
1 for FAIL
Example:
//Perform a measurement and query the result of the check for
limit line 1.
CALC:LIM:UPP:THR -70
INIT;*WAI
CALC:LIM1:FAIL?
Usage:
Query only
CALCulate<n>:LIMit<k>:LOWer:SELect
<LimitLine>
This command selects the lower limit line.
You can either select a limit line that is already stored on the internal memory or create
.
Suffix:
<n>
.
irrelevant
<k>
Setting parameters:
<LimitLine>
String containing the file name of the lower limit line.
Example:
//Select the lower limit line
FREQ:SPAN 0HZ
CALC:LIM:LOW:SEL 'GSM_Lower.rellim'
Usage:
Setting only
CALCulate<n>:LIMit<k>:LOWer:THReshold
<Level>
This command defines the level of a lower threshold limit line.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Level>
<numeric value>
Unit depends on the unit you have currently selected for the ver-
tical axis.
Example:
//Turn on threshold limit line
CALC:LIM:LOW:THR -40DBM
CALCulate<n>:LIMit<k>:STATe
<State>
This command turns a limit check on and off.
You can query the result of the limit check with
Remote Commands of the Spectrum Application