Remote Control
FPC1000
193
User Manual 1178.4130.02 ─ 02
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<ReferenceLevel>
<numeric value>
The unit depends on
The available value range is specified in the data sheet.
*RST:
-20 dBm
Example:
//Define reference level
UNIT:POW DBM
DISP:TRAC:Y:RLEV –60
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Offset>
<numeric value>
Range:
-100 to 100 dB
*RST:
0 dB
Default unit: dB
Example:
//Define a level offset
DISP:TRAC:Y:RLEV:OFFS –10dB
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the position of the reference level on the display grid.
Prerequisites for this command
●
Select logarithmic scale for the y-axis (
DISPlay<n>[:WINDow]:TRACe<t>:Y:
For a linear scale, you cannot modify the reference position.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Remote Commands of the Spectrum Application