![Rohde & Schwarz FPC1000 User Manual Download Page 212](http://html.mh-extra.com/html/rohde-and-schwarz/fpc1000/fpc1000_user-manual_1477825212.webp)
Remote Control
FPC1000
197
User Manual 1178.4130.02 ─ 02
Prerequisites for this command
●
[SENSe:]CORRection:TRANsducer<td>:SELect
).
Suffix:
<td>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF | 0
Example:
//Turn on primary transducer
CORR:TRAN1:SEL 'HE200A-20-200.pritrd'
CORR:TRAN1 ON
//Turn on secondary transducer
CORR:TRAN2:SEL 'PreAmp.sectrd'
CORR:TRAN2 ON
UNIT<n>:POWer
<Unit>
This command selects the unit displayed on the y-axis.
Suffix:
<n>
.
irrelevant
Parameters:
<Unit>
DBM | DBUV | DBMV | V | W | DUVM | DUAM | V_M | W_M2
Note that the availability of units depends on the operating
mode.
*RST:
dBm
Example:
//Select displayed unit
UNIT:POW DBUV
13.10.4 Bandwidth
.........................................................................................197
[SENSe:]BANDwidth[:RESolution]
................................................................................... 197
[SENSe:]BWIDth[:RESolution]:AUTO
...............................................................................198
[SENSe:]BANDwidth[:RESolution]:AUTO
..........................................................................198
............................................................................................ 198
........................................................................................199
...................................................................................199
[SENSe:]BWIDth[:RESolution]
<Bandwidth>
[SENSe:]BANDwidth[:RESolution]
<Bandwidth>
This command selects the resolution bandwidth.
Remote Commands of the Spectrum Application
北京海洋兴业科技股份有限公司(证券代码:839145)
www.hyxyyq.com