Remote Control
FPC1000
240
User Manual 1178.4130.02 ─ 02
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:ADEMod:THD[:RESult]?
This command queries the THD result.
Suffix:
<n>
.
irrelevant
Return values:
<THD>
<numeric value>
Default unit: dB
Example:
//Query THD result
CALC:MARK:FUNC:ADEM:THD?
Usage:
Query only
[SENSe:]ADEMod:FM:OFFSet?
This command queries the frequency deviation.
Prerequisites for this command
●
).
Return values:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Query frequency deviation
CALC:FEED 'XTIM:FMS'
ADEM:FM:OFFS?
Usage:
Query only
TRACe[:DATA]?
This command queries the trace data (measurement results in graphical result dis-
plays).
Prerequisites for this command
●
Select graphical result display (
).
Return values:
<Results>
Returns one value for each measurement point.
The type of value depends on the measurement (Hz for FM,
PCT for AM).
Example:
//Query FM results
CALC:FEED 'XTIMe:FM'
TRAC?
Usage:
Query only
Remote Commands of the Analog Demodulation
北京海洋兴业科技股份有限公司(证券代码:839145)