Remote Control
FPC1000
201
User Manual 1178.4130.02 ─ 02
Parameters:
<State>
ON | 1
Select continuous sweep mode.
OFF | 0
Select single sweep mode.
Example:
//Turn on single sweep mode.
INIT:CONT OFF
INITiate[:IMMediate]
The command initiates a new measurement sequence.
Effects of this command
●
In single sweep mode, with an average count > 0, the command restarts the mea-
surement.
●
When you have selected the maxhold, minhold or average trace mode the com-
mand resets the trace.
in single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Example:
//Start a single sweep and wait to end of sweep
INIT:CONT OFF
INIT;*WAI
Usage:
Event
13.10.6 Trigger
TRIGger[:SEQuence]:HOLDoff[:TIME]
..............................................................................201
TRIGger[:SEQuence]:LEVel:VIDeo
..................................................................................202
.......................................................................................... 202
........................................................................................202
TRIGger[:SEQuence]:HOLDoff[:TIME]
<TriggerDelay>
This command defines the length of the trigger delay.
Prerequisites for this command
●
Select external trigger (
Parameters:
<TriggerDelay>
<numeric value>
Range:
0 to 100 s
*RST:
0 s
Default unit: s
Remote Commands of the Spectrum Application