Remote Control
FPC1000
239
User Manual 1178.4130.02 ─ 02
Query parameters:
<MeasType>
MIDDle
Queries the average frequency deviation.
MPEak
Queries the minimum frequency deviation.
PPEak
Queries the maximum frequency deviation.
RMS
Queries the RMS value of the frequency deviation.
Return values:
<Result>
Default unit: Hz
Example:
//Query average frequency deviation
CALC:FEED 'XTIM:FMS'
CALC:MARK:FUNC:ADEM:FM? MIDD
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:ADEMod:MINDex[:RESult]?
This command queries the modulation index.
Prerequisites for this command
●
).
Suffix:
<n>
.
irrelevant
Return values:
<ModIndex>
<numeric value>
Default unit: PCT
Example:
//Query modulation index
CALC:FEED 'XTIM:AMS'
CALC:MARK:FUNC:ADEM:MIND?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:ADEMod:SINad[:RESult]?
This command queries the SINAD result.
Suffix:
<n>
.
irrelevant
Return values:
<SINAD>
<numeric value>
Default unit: dB
Example:
//Query SINAD result
CALC:MARK:FUNC:SIN?
Remote Commands of the Analog Demodulation
北京海洋兴业科技股份有限公司(证券代码:839145)