Remote Control
FPC1000
248
User Manual 1178.4130.02 ─ 02
Suffix:
<n>
.
irrelevant
Return values:
<ModIndex>
<numeric value>
Default unit: PCT
Example:
//Query modulation index
CALC:MARK:FUNC:DDEM:MIND?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MOERror[:RESult]?
This command queries the modulation error.
Suffix:
<n>
.
irrelevant
Return values:
<ModError>
<numeric value>
Default unit: PCT
Example:
//Query modulation error
CALC:MARK:FUNC:DDEM:MOER?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MPOWer[:RESult]?
This command queries the power of the carrier signal.
Suffix:
<n>
.
irrelevant
Return values:
<Power>
<numeric value>
Default unit: dBm
Example:
//Query carrier power
CALC:MARK:FUNC:DDEM:MPOW?
Usage:
Query only
[SENSe:]DDEMod
<Measurement>
This command selects the demodulation scheme of the signal you are measuring.
Remote Commands of the Digital Demodulation