The Receiver Application
FPC1000
117
User Manual 1178.4130.02 ─ 02
3. Select the "Channel Downlink" or "Channel Uplink" menu item to select a channel
table.
10.3 Amplitude Configuration
Access
: "Ampt"
Access
: "Setup" > "Config Overview" > "Amplitude" (selected settings)
Access
: "Setup" > "Config Overview" > "Input" (selected settings)
Remote commands to configure the amplitude:
●
The following amplitude settings available in the receiver application are the same as
in the spectrum application.
●
"Reference Level":
Chapter 9.4.1, "Reference Level"
●
"Range / Ref Position" > "Auto Range":
Chapter 9.4.3, "Scale of the Y-Axis"
●
"Range / Ref Position" > "Range <x>dB":
Chapter 9.4.3, "Scale of the Y-Axis"
●
"Range / Ref Position" > "Linear 0 - 100%":
Chapter 9.4.3, "Scale of the Y-Axis"
●
"Range / Ref Position" > "Ref Position":
"Defining the position of the reference
●
"Range / Ref Position" > "Reference Offset":
●
"Unit":
●
"Preamplifier":
●
"Attenuator":
●
"Impedance":
"Selecting the input impedance"
●
"Transducer":
Defining the bargraph scale
The scale of the bargraph depends on the
and the level range.
The reference level is the maximum power level that the bargraph displays. You should
set the reference level in a way that the signal level does not exceed the reference
level. It should also be high enough so that the signal does not disappear in the inher-
ent noise.
1. Press the "Ampt" key to open the amplitude menu.
2. Select the "Reference Level" menu item.
3. Define the reference level that best fits the measurement.
If you measure unknown signals, define a high reference level in combination with
a high attenuation to avoid an input overload.
Amplitude Configuration