![Rohde & Schwarz FPC1000 User Manual Download Page 185](http://html.mh-extra.com/html/rohde-and-schwarz/fpc1000/fpc1000_user-manual_1477825185.webp)
Remote Control
FPC1000
170
User Manual 1178.4130.02 ─ 02
SYSTem:BEEPer:KEY:VOLume
<Volume>
This command defines the volume of the keyboard click sound.
Parameters:
<Volume>
<numeric value>
Range:
0 to 1
*RST:
0.3 (30%)
Example:
//Select the key click volume
SYST:BEEP:KEY:VOL 0.10
SYSTem:BEEPer:POVerload[:STATe]
<State>
This command turns an alert on a power overload on and off.
You can define the volume of this sound with
Parameters:
<State>
ON | 1
An overload situation on one of the inputs produces a sound.
OFF | 0
An overload produces no sound.
*RST:
OFF
Example:
//Turn on audio output in case of overload
SYST:BEEP:POV ON
SYSTem:BEEPer:VOLume
<Volume>
This command defines the volume of the system beeper.
Parameters:
<Volume>
<numeric value>
Range:
0 to 1
*RST:
0.6 (60%)
Example:
//Define the volume of the system beeper
SYST:BEEP:VOL 0.50