Remote Control
FPC1000
192
User Manual 1178.4130.02 ─ 02
Parameters:
<Scaling>
LOGarithmic
Select a logarithmic scale.
When you select a logarithmic scale, you can define the display
range manually or automatically.
•
Manual:
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]
•
DISPlay<n>[:WINDow]:TRACe<t>:Y[:
LINear
Select a linear scale (%).
The display range of a linear scale is fixed (0
% to 100
%).
*RST:
LOGarithmic
Example:
//Select linear scaling of the y-axis
DISP:TRAC:Y:SPAC LIN
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]
<DisplayRange>
This command defines the display range of the vertical axis.
Prerequisites for this command
●
Select logarithmic scale for the y-axis (
DISPlay<n>[:WINDow]:TRACe<t>:Y:
For a linear scale, you cannot modify the display range as it is fixed.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<DisplayRange>
<numeric value>
Range:
1 to 150 dB
*RST:
100 dB
Default unit: dB
Example:
//Define display range
DISP:TRAC:Y:SPAC LOG
DISP:TRAC:Y 110dB
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:ADJust
This command automatically scales the y-axis.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Example:
//Adjust the y-axis
DISP:TRAC:Y:ADJ