The Receiver Application
FPC1000
119
User Manual 1178.4130.02 ─ 02
● "CISPR BW: Auto": Automatically selects a 6 dB bandwidth, depending on the
receiver frequency (see table above).
10.5 Sweep Configuration
Access
: "Sweep"
Access
: "Setup" > "Config Overview" > "Bandwidth" (selected settings)
Remote commands to control the measurement:
●
Chapter 13.11.5, "Sweep Control"
The sweep menu provides settings to control the measurement and the way data is
captured.
Defining the measurement time
The measurement time is the time that the R&S
FPC1000 collects data at each mea-
surement frequency to calculate the results for that frequency according to the detector
you have selected.
1. Press the "Sweep" key to open the sweep menu.
2. Select the "Measurement Time" menu item.
3. Define the measurement time you need.
Selecting the measurement mode
When you enter the receiver mode, the R&S
FPC1000 repeatedly measures the
receive frequency or set of frequencies over the measurement time you have defined.
If you want to measure a fixed frequency or frequency scan only once, select single
scan mode.
1. Press the "Sweep" key to open the sweep menu.
2. Select the "Continuous Scan" menu item.
The R&S
FPC1000 starts a continuous measurement of the selected
You can interrupt a continuous measurement by selecting the "Continuous Sweep"
menu item again.
3. Select the "Single Scan" menu item.
The R&S
FPC1000 starts a single measurement of the selected
.
Sweep Configuration