Remote Control
FPC1000
242
User Manual 1178.4130.02 ─ 02
13.12.4 Bandwidth
[SENSe:]ADEMod:BWIDth:DEModulation
.........................................................................242
[SENSe:]ADEMod:BANDwidth:DEModulation
................................................................... 242
[SENSe:]FILTer:DEMPhasis[:STATe]
...............................................................................242
[SENSe:]FILTer:DEMPhasis:TCONstant
.......................................................................... 242
[SENSe:]ADEMod:BWIDth:DEModulation
<Bandwidth>
[SENSe:]ADEMod:BANDwidth:DEModulation
<Bandwidth>
This command selects the demodulation bandwidth.
Parameters:
<Bandwidth>
<numeric value> (integer only)
Refer to the datasheet for a list of supported bandwidths.
*RST:
200 kHz
Default unit: Hz
Example:
//Select demodulation bandwidth
ADEM:BAND:DEM 50KHZ
[SENSe:]FILTer:DEMPhasis[:STATe]
<State>
This command turns the deemphasis filter on and off.
You can select the time constant of the filter with
.
Prerequisites for this command
●
Select 200 kHz or 300 kHz demodulation bandwidth (
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turn on deemphasis filter
ADEM:BAND:DEM 300KHZ
FILT:DEMP ON
[SENSe:]FILTer:DEMPhasis:TCONstant
<Time>
This command selects the time constant of the deemphasis filter.
Prerequisites for this command
●
Select 200 kHz or 300 kHz demodulation bandwidth (
●
[SENSe:]FILTer:DEMPhasis[:STATe]
).
Remote Commands of the Analog Demodulation