Remote Control
FPC1000
246
User Manual 1178.4130.02 ─ 02
Return values:
<Power>
<numeric value>
Default unit: dBm
Example:
//Query carrier power
CALC:MARK:FUNC:DDEM:CARR?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:CFDRift[:RESult]?
This command queries the carrier frequency drift of the measured signal.
Prerequisites for this command
●
).
Suffix:
<n>
.
irrelevant
Return values:
<Frequency>
<numeric value>
Default unit: Hz/symbol
Example:
//Query carrier frequency drift
CALC:MARK:FUNC:DDEM:CFDR?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:CFERror[:RESult]?
This command queries the carrier frequency error of the measured signal.
Suffix:
<n>
.
irrelevant
Return values:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Query carrier frequency error
CALC:MARK:FUNC:DDEM:CFER?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MDEPth[:RESult]?
This command queries the modulation depth of the measured signal.
Prerequisites for this command
●
Select ASK demodulation (
Suffix:
<n>
.
irrelevant
Remote Commands of the Digital Demodulation