Remote Control
FPC1000
194
User Manual 1178.4130.02 ─ 02
Parameters:
<Position>
<numeric value> (integer)
Range:
0 to 10
*RST:
10
Example:
//Set the reference position to the 5th grid line
DISP:TRAC:Y:SPAC LOG
DISP:TRAC:Y:RPOS 5
INPut:ATTenuation
<Attenuation>
This command selects the RF attenuation.
Effects of this command
●
Turns off automatic selection of the attenuation and therefore the coupling to the
).
●
Reference level will be adjusted if the currently defined reference level is not com-
patible to the selected attenuation.
Parameters:
<Attenuation>
<numeric value>
Range:
0 to 40
Increment: 5
*RST:
0 (AUTO is ON)
Default unit: dB
Example:
//Select attenuation
INP:ATT 30dB
INPut:ATTenuation:AUTO
<State>
This command turns automatic selection of the attenuation level on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON | 1
Example:
//Turn on automatic attenuation selection
INP:ATT:AUTO ON
INPut:ATTenuation:MODE
<AttenuationMode>
This command selects the attenuation mode.
Prerequisites for this command
●
Select attenuation automatically (
Remote Commands of the Spectrum Application