
Remote Control
FPC1000
207
User Manual 1178.4130.02 ─ 02
[SENSe:]DETector<t>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
Suffix:
<t>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON | 1
Example:
//Turn off automatic detector selection
DET:AUTO OFF
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of averages for the average trace mode.
The R&S
FPC1000 performs one sweep for sweep count 0 or 1.
Prerequisites for this command
●
DISPlay<n>[:WINDow]:TRACe<t>:MODE
).
Parameters:
<SweepCount>
<numeric value> (integer)
Range:
0 to 999
*RST:
1
Example:
//Average over 64 sweeps
DISP:TRAC:MODE AVER
SWE:COUN 64
//Turn on single sweep mode, start the sweep and wait for its
end
INIT:CONT OFF
INIT;*WAI
TRACe<t>[:DATA]?
[<ResultType>]
This command queries the trace data of the current measurement.
It also transfers data from a file to a particular trace.
With the command
, you can set the data format.
Suffix:
<t>
.
Parameters:
<ResultType>
TRACE1
Query the data of trace 1.
TRACE2
Query the data of trace 2.
Remote Commands of the Spectrum Application
北京海洋兴业科技股份有限公司(证券代码:839145)
www.hyxyyq.com